Anaesthesia Workstation is part of the Airway Management and Resuscitation section of Anaesthesia. Practise 41 exam-style MCQs for NEET PG, FMGE, INI-CET and NEET SS, each with the correct answer and a detailed explanation.
Anaesthesia Workstation MCQs inspect vacuum-insulated evaporators, vaporizer ideals, cylinder testing, pressure circuits and fail-safe valves against hypoxic mixtures. These high-yield NEET PG questions on the machine that delivers anaesthesia.
Sample Anaesthesia Workstation MCQs with answers
Sample MCQ #1
When compared to a cylinder manifold, a vacuum-insulated evaporator (VIE) is the most economical way to store and supply oxygen in a large hospital. What is the single most important reason for this?
AThe critical temperature of liquid oxygen is ideal to store as a liquid
BLiquid oxygen is easier to store.
CVIE does not require rigorous maintenance.
DWhen evaporated, liquid oxygen occupies a large volume as gas.
Show Answer & Explanation ▾
✓ Correct Answer: D. When evaporated, liquid oxygen occupies a large volume as gas.
Liquid oxygen is stored in a vacuum-insulated evaporator (VIE) at a temperature of -150°C to -170°C and at a pressure of 5-10 atmospheres. In simple terms this works like a giant thermos flask. At 15°C, one volume of liquid oxygen occupies 842 times its volume as gas. Oxygen constantly evaporates from the top of the VIE. As this oxygen is very cold, it needs to be passed through a super heater. When the oxygen flows at a faster rate, the temperature of the liquid falls due to the latent heat of evaporation. To overcome this problem additional heat is provided by a pressure-raising vaporiser. Regular maintenance, monitoring of usage of oxygen and topping up the tank at regular intervals is necessary. Reference: Al-Shaikh B, Stacey S, Eds. Liquid oxygen. In: Essentials of anaesthetic equipment, 3rd ed. London, UK: Churchill Livingstone, Elsevier, 2007; Chapter 1: 9-10.
Sample MCQ #2
True statement about Volatility and vaporization?
AThe more volatile a liquid, the faster it vaporizes
BVolatility and vaporization are not related
CThe more volatile a liquid, the slower it vaporizes
DNone
Show Answer & Explanation ▾
✓ Correct Answer: A. The more volatile a liquid, the faster it vaporizes
The more volatile a liquid, the faster it vaporizes and the greater the loss of latent heat of vaporization Liquid cools à rate of vaporization falls Reference : Anaesthesia Review a Comprehensive Texbook 5th Edition Chapter 44 Page No .423
Sample MCQ #3
The ideal vaporizer should be?
AEasy to transport
BCorrosion and solvent resistant
CLightweight, robust and durable
DAll of the above
Show Answer & Explanation ▾
✓ Correct Answer: D. All of the above
The ideal vaporizer should be Lightweight, robust and durable Easy to transport Corrosion and solvent resistant Leak proof Economical and safe to use Require minimal servicing Accurate over a wide range of: Freah Gas Flow Liquid agent levels Ambient pressures Ambient temperatures Unaffected by: Heat loss due to vaporization Pressure changes downstream of the vaporizer Tilting or tipping Compatible with vaporizer interlock systems Reference :Anaesthesia Review a Comprehensive Texbook 5th Edition Chapter 44 Page No .423 to 428
Sample MCQ #4
Which test is performed on gas cylinders to check the quality of manufacturing?
AThe tensile test
BFlattening test
CBend test
DAll the above
Show Answer & Explanation ▾
✓ Correct Answer: D. All the above
The cylinders used in medical practice are either solid drawn or made from seamless steel tube, and never welded at the ends. After manufacturing, the cylinders are kept in a furnace at a temperature of 860-890 degree C and then carefully removed and cooled at atmospheric pressure . As the cylinders must withstand a high gas pressure and rough handling during transport, they are subjected to mechanical testing before marketing for use. The tensile test, flattening test, impact test and bend test are performed on one cylinder from a batch of every 100 finished cylinders. Reference: Anaesthesia Review: A Comprehensive Textbook by Atul, 5th Edition, Page Number 396-409.
Sample MCQ #5
1 kPa is approximately equivalent to:
A10 cm H2O
B14.7 psi
C0.76 mm Hg
D100 mbar
Show Answer & Explanation ▾
✓ Correct Answer: A. 10 cm H2O
Explanation: Pressure Conversions: 1 kPa = 10 cm H2O is a valid approximation, which is useful in clinical settings for measuring pressures in respiratory physiology and anaesthesia. To understand this, atmospheric pressure is often measured in various units, including mmHg, mbar, and psi. Atmospheric Pressure Reference: Atmospheric pressure is defined as the pressure exerted by the weight of air in the atmosphere, commonly measured using a barometer. 760 mmHg (Torr) is equivalent to 1 bar = 1000 mbar = 101.3 kPa . Conversions for Context: 1 kPa ≈ 7.5 mmHg (Option C) 1 bar = 14.5 psi , thus 1 kPa ≈ 0.145 psi (Option B) 1 kPa is equivalent to 10 mbar , not 100 mbar (Option D) Key Takeaways: 1 kPa is approximately 10 cm H2O, a key conversion for understanding pressures in clinical practice. Familiarity with pressure units is essential in anaesthesia and critical care settings.
Sample MCQ #6
Which of the following is NOT found in the low pressure circuit on an anesthesia machine?
AOxygen supply failure alarm
BFlowmeters
CVaporizers
DVaporizer check valve
Show Answer & Explanation ▾
✓ Correct Answer: A. Oxygen supply failure alarm
The anesthesia machine, now more properly called the anesthesia workstation, has two main pressure circuits. The higher-pressure circuits consist of the gas supply from the pipelines and tanks, all piping, pressure gauges, pressure reduction regulators, check valves (which prevent backward gas flow), the oxygen pressure sensor shutoff valve (also called the oxygen failure cutoff or fail-safe valve), the oxygen supply failure alarm, and the oxygen flush valve-or, simplistically, everything up to the gas flow control valves and the machine common gas outlet. The low-pressure circuit starts with and includes the gas flow control valves, flowmeters, vaporizers, and vaporizer check valve and goes to the machine common gas outlet. See also figure for explanation to Question 12 Reference: Barash: Clinical Anesthesia, ed 7, pp 641-650; Miller: Basics of Anesthesia, ed 6, pp 198-204.
Sample MCQ #7
Which of the following prevents delivery of hypoxic gas mixture once the oxygen pressure falls below 25 psi?
ADiameter index safety system
BPin index safety system
CInspiratory check valve
DFail-safe valve
Show Answer & Explanation ▾
✓ Correct Answer: D. Fail-safe valve
The fail-safe valve automatically closes nitrous oxide (and other gases) to prevent delivery of hypoxic gas mixture to the patient. The fail-safe valve is designed to be activated when oxygen pressure falls below 25 psi. Reference: Lippincott's ANESTHESIA REVIEW: 1,001 QUESTIONS AND ANSWERS Paul Sikka, MD, PhD
Sample MCQ #8
The oxygen-flush valve provides which of the following oxygen flows (L/min) to the common gas outlet?
A10
B25
C50
D90
Show Answer & Explanation ▾
✓ Correct Answer: C. 50
The oxygen-flush valve provides gas flow at pipeline pressures of about 45 to 55 psi at 35 to 75 L/min. The high flow of oxygen is provided directly to the common gas outlet, bypassing the flowmeters and vaporizers. One should be careful when using the oxygen-flush valve, as high gas flows at high pressures can cause lung barotrauma in the patient. Reference: Lippincott's ANESTHESIA REVIEW: 1,001 QUESTIONS AND ANSWERS Paul Sikka, MD, PhD
What you can practise in Anaesthesia Workstation
41 Anaesthesia Workstation MCQs on Airway Management and Resuscitation — each with detailed explanations
Previous-year question practice across NEET PG, FMGE, INI-CET and NEET SS
Progress and performance tracking to measure retention
Anaesthesia Workstation MCQs inspect vacuum-insulated evaporators, vaporizer ideals, cylinder testing, pressure circuits and fail-safe valves against hypoxic mixtures. These high-yield NEET PG questions on the machine that delivers anaesthesia.
How many Anaesthesia Workstation questions are there, and which exams do they cover?
MedLadder includes 41 curated MCQs for Anaesthesia Workstation under Anaesthesia (Airway Management and Resuscitation), covering NEET PG, FMGE, INI-CET and NEET SS preparation.